Enable EMI mode with SSA3021X SSA3032X models
January 22, 2021
On SSA3021X and SSA3032X models, you can enable the EMI specific settings by:
- Activate the EMI filter by pressing BW > Set Filter to EMI
-The Gaussian filter has a filter shape defined at 3 dB. The EMI filter has it’s shape defined at 6 db.
2. When the EMI filter is selected, the EMI RBW settings are unlocked. These are 250 Hz, 9 kHz, and 120 kHz.
3. You can also enable the QuasiPeak detector by pressing Detect > and select Quasi Peak
More information on spectrum analyzer bandwidth can be found here: Spectrum Analyzer Bandwidth