Do your instruments support UDP communications over LAN?

August 5, 2021

Not at this time. Many instruments support communication via sockets (port 5025) and telnet (port 5024) but the port numbers cannot be changed. If you use VISA, you can implement LAN via TCP/IP, which does not need a port number. Here is a listing of current SIGLENT instruments with Socket and Telnet Support.

SDG available waveform lengths

July 29, 2021

The SDG series of arbitrary waveform generators allow you to create any function you wish. Here are the memory depths available for building arbs for each model:

How to check available internal memory?

July 29, 2021

Simply access the Save/Recall menu for the instrument in question and select Local. The menu will show used/total amounts. Here, the SDG6X has approximately 83 MB and the SDG1X has 85 MB of memory:

Are SIGLENT products certified by a Nationally Recognized Test Laboratories (NRTL)?

May 7, 2021

Per OSHA, a Nationally Recognized Testing Laboratory (NRTL) is an independent third-part laboratory recognized by the United States Occupational Safety and Health Administration (OSHA) to test and certify products to applicable product safety standards. Examples of NRTLs include UL, CSA, and TUV. Many SIGLENT products are TUV certified, which also covers UL and CSA requirements. To … Continued

How do I directly control an instrument using LAN?

February 16, 2021

For convenience, we recommend connecting your instrument to your LAN using a router that can configure DHCP directly and use an internet switch to connect the LAN to the instruments you wish to control. In some cases, you may wish to directly connect your computer to an instrument LAN port without a LAN available. This … Continued

Why does my function generator amplitude not match the set point?

March 3, 2020

The most likely cause is that the generator output is not configured with the correct value that matches the actual load. Function generators like the SIGLENT SDG series are designed to have an output impedance of 50 ohms. This output impedance is assumed to be in series with a known load impedance R2, as shown … Continued

Currently supported instruments for use with the SIGLENT USB-GPIB adapter

August 8, 2019

The SIGLENT USB-GPIB adaptor currently works with the following SIGLENT instruments: Multimeters – SDM3065X/SDM3065X-SC SDM3055/SDM3055-SC SDM3045X   Function/Arbitrary waveform generators – SDG6000X Series SDG5000 Series SDG2000X Series SDG1000X Series SDG1000 Series   Spectrum & vector network analyzers – SSA3000X Series SSA3000X Plus Series SVA1000X Series   Oscilloscopes – SDS1000CFL 

SIGLENT SPA1010 Amplifier output control

January 15, 2019

The SIGLENT SPA1010 Amplifier can be a useful accessory for applications requiring higher voltage and/or current from a signal source like a SIGLENT SDG arbitrary waveform generator. The SPA features a power-on indicator LED, Short LED, a X10 indicator LED, and a set of switches on the front panel. To use: Connect the output of … Continued