Instrument Socket and Telnet Port Information

February 25, 2021

Product Type Product Series Socket port 5025 (Y/N) Telnet port 5024(Y/N) Digital Oscilloscope SDS5000X Y Y SDS2000X Plus Y Y SDS2000X N N SDS2000X-E Y Y SDS1000X N N SDS1000X-E Y Y SDS1104X-U Y Y SDS1000CML+ N N SDS1000DL+ N N Waveform Generator SDG6000X Y Y SDG2000X Y Y SDG1000X Y Y SDG800 N N … Continued

How do I directly control an instrument using LAN?

February 16, 2021

For convenience, we recommend connecting your instrument to your LAN using a router that can configure DHCP directly and use an internet switch to connect the LAN to the instruments you wish to control. In some cases, you may wish to directly connect your computer to an instrument LAN port without a LAN available. This … Continued

Updated Frequency Uncertainty Calculations (SSA, SSA Plus, SVA)

August 26, 2020

Currently listed datasheet calculation for SSA Frequency Uncertainty is being updated, but as of 08/26/2020, it has not.  Here is the new calculation: We can use below new specification: ± (frequency indication × frequency reference uncertainty + 1/2(span/(sweep points-1)) + 10% × resolution bandwidth + marker resolution)   For example: The frequency indication is 500 … Continued

Which values do I enter for antenna correction factors on my spectrum analyzer?

December 27, 2019

If you have an antenna with calibration data, chances are it includes a number of columns of data. One column should be the frequency being tested. The others can include free-range performance, gain, and more. For reference, Antenna Factor (or correction factor) is defined as the ratio of the incident Electromagnetic Field to the output voltage … Continued

SVA1000X series Port 1 input damage limits for reflected/S11/VSWR measurements

December 12, 2019

Port 1 of the SIGLENT SVA1000X Series of Vector Network Analyzers is used as the output for the tracking generator as well as the S11/Reflected energy port. The damage threshold is approximately +20 dBm and this port does not feature any “over-limit” warning. When measuring VSWR/Reflected energy/S11 on antennas or devices that may be able … Continued