How to view multiple traces without all of them using the QuasiPeak Detector setting
March 19, 2020
The SIGLENT SSA3000X, SSA3000X Plus, and SVA1000X products all have an EMI option that adds a Quasi- Peak detector selection. The Quasi Peak detector is an integrating detector type that will automatically apply to all active traces on the analyzer. To prevent existing traces from being overwritten with the Quasi Peak detector data, please follow … Continued
Which values do I enter for antenna correction factors on my spectrum analyzer?
December 27, 2019
If you have an antenna with calibration data, chances are it includes a number of columns of data. One column should be the frequency being tested. The others can include free-range performance, gain, and more. For reference, Antenna Factor (or correction factor) is defined as the ratio of the incident Electromagnetic Field to the output voltage … Continued
SVA1000X series Port 1 input damage limits for reflected/S11/VSWR measurements
December 12, 2019
Port 1 of the SIGLENT SVA1000X Series of Vector Network Analyzers is used as the output for the tracking generator as well as the S11/Reflected energy port. The damage threshold is approximately +20 dBm and this port does not feature any “over-limit” warning. When measuring VSWR/Reflected energy/S11 on antennas or devices that may be able … Continued
SVA Options: How do I know what is activated/available?
October 1, 2019
The SIGLENT SVA series of analyzer instruments are very flexible platforms that offer a number of different functions that can be activated. New instruments are delivered with over 100 hours of trial time for each option, allowing users to try the options before purchasing them. The trial timer is running while the instrument is powered … Continued
How to configure the SVA1X for complex impedance measurements
August 21, 2019
The SIGLENT SVA1X series is a multi-purpose RF measurement instrument. This operating tip shows you how to quickly configure the Vector Network mode to show a Smith chart displaying complex impedance: VNA mode – Press Mode > VNA Set Meas type as S11 or S21 Set Format – Smith Select R + … Continued
Currently supported instruments for use with the SIGLENT USB-GPIB adapter
August 8, 2019
The SIGLENT USB-GPIB adaptor currently works with the following SIGLENT instruments: Multimeters – SDM3065X/SDM3065X-SC SDM3055/SDM3055-SC SDM3045X Function/Arbitrary waveform generators – SDG6000X Series SDG5000 Series SDG2000X Series SDG1000X Series SDG1000 Series RF Generators – SSG3000X/-IQE Series SSG5000X/-V Series Spectrum & vector network analyzers – SSA3000X Series SSA3000X Plus Series SVA1000X Series Oscilloscopes … Continued
USB host connector power limits for instrumentation
June 14, 2019
SIGLENT instruments with USB ports can supply 500mA steady current at 5 V.
EasySpectrum: Save/Recall State files
March 7, 2019
EasySpectrum is free software used to control and collect data from SIGLENT SSA and SVA spectrum analyzer products. State files are used by the SIGLENT SSA and SVA instruments to recall or save specific instrument configuration settings. By storing and recalling states, you can quickly configure the analyzer for different tasks. To use State files … Continued
Where do I find environmental, EMC, and other compliance information for your products?
January 10, 2019
Information regarding product compliance is available on the datasheet for each product. Contact your local Siglent office for more details if you have trouble finding what you need. You can also write support@siglentna.com
Correction factors in Monitor Mode with EasySpectrum software
December 14, 2018
SIGLENT EasySpectrum software includes the ability to create and edit correction factors that can be useful in correcting cable and adapter losses as well as adjusting spectral data for antenna gain. To create a correction file: Download EasySpectrum. Here is a link to the EasySpectrum Getting Started Guide Connect your SIGLENT analyzer to the controlling computer … Continued